ACCULINK
It is a pre-mounted self-expanding nickel-titanium (Nitinol) stent with open cell
design and high adaptability to the arterial morphology.
A minimum foreshortening of the stent is reported.
The delivery system (132cm long) has an internal hypotube and works on a rapid
exchange system, 0.014" compatible, with a low profile (5.9 Fr).
A minimum 6Fr introducer sheath or a 8 Fr guiding catheter, no longer than 100cm,
is required.
Two radiopaque markers are placed on the shaft to mark the stent location.
The stent is deployed using a pullback handle; once deployment is initiated the
stent cannot be repositioned or recaptured.
To avoid complications, it is advisable to re-advance the sheath and relock the
delivery system before removal of the sheath or of the guiding catheter.
Both cylindrical and tapered configurations are available to fit different arterial
morphologies.
RX ACCULINK Carotid Cylindrical Stent Reference Vessel Chart
|
STENT DIAMETER mm |
STENT LENGTH mm |
REFERENCE VESSEL mm |
|
5.0 |
20-30-40 |
3.6-4.5 |
|
6.0 |
20-30-40 |
4.3-5.4 |
|
7.0 |
20-30-40 |
5.0-6.4 |
|
8.0 |
20-30-40 |
5.7-7.3 |
|
9.0 |
20-30-40 |
6.4-8.2 |
|
10.0 |
20-30-40 |
7.1-9.1 |
RX ACCULINK Carotid Tapered Stent Reference Vessel Chart
|
STENT DIAMETER mm |
STENT LENGTH mm |
DISTAL REFERENCE VESSEL mm |
PROXIMAL REFERENCE VESSEL mm |
|
6-8 |
30-40 |
4.3-5.4 |
5.7-7.3 |
|
7-10 |
30-40 |
5.0-6.4 |
7.1-9.1 |
 |