PRECISE PRO RX


It is a self-expanding, nitinol (nickel-titanium alloy) stent extremely flexible composed of fine mesh tubular prosthesis, which achieves its unconstrained diameter upon deployment into the carotid arteries.
An open cell design grants a high flexibility to this stent that proves to be fit even for extremely tortuous vessels.
In addition this stent can be considered "self-tapered" because its design makes it particularly suitable for patients whose CCA and ICA calibres are different.
The New Precise Pro Rx is constrained within the space between the wire lumen and the distal section of the outer sheath.
The stent is deployed with a "pin & pull" mechanism by retrieving the outer sheath.
Two markers, at the proximal and distal end of the Precise Pro, allow a correct deployment of the stent that is achieved when the outer sheath radiopaque marker is proximal to the inner shaft marker.
The stent shaft presents a very low-profile with high pushability and a unique coil inner shaft design avoids any stent movement during the releasing phase A 5 Fr introducer sheath or a 7 Fr guiding catheter is required when a 5 to 8 mm Precise Pro stent is used.
Instead for the 9 and 10 mm Precise Pro a 6 Fr introducer or an 8 Fr guiding catheter is needed.
The stent is MRI safe and does not interfere with the operation of an MRI device with no imaging artefacts.

PRECISE Carotid Reference Vessel Chart
STENT DIAMETER mm STENT LENGTH mm REFERENCE VESSEL mm
5.0 * 20-30-40 3.0-4.0
6.0 * 20-30-40 4.0-5.0
7.0 * 20-30-40 5.0-6.0
8.0 * 20-30-40 6.0-7.0
9.0 ** 20-30-40 7.0-8.0
10.0 ** 20-30-40 8.0-9.0
* 5 Fr (2.0 mm) compatible delivery system
** 6 Fr (2.3 mm) compatible delivery system